Table 4-8. Command and Query Reference (cont)

MEASUREMENT

UNITS STRING

Scaled “MX+B”
Volts DC

“VDC”

Volts AC

“VAC”

Resistance

“OHMS”

Frequency

“Hz”

Temperature

“C”

Temperature

“F”

FORMAT?

Query Response Format

Returns the format presently in use:

The default. IEEE-488.2-compatible. No measurement units are allowed.

Responses are RS-232-compatible, allowing a means of recording the
units with the measurement value.

FUNC

Channel Function Definition

Define the measurement function and range for the indicated channel.
Attempting to use the FUNC command when measurements are active
results in an Execution Error.

Note

Successful execution of the FUNC command clears any alarm limits and
scaling values for this channel. Therefore, you must define a channel’s
function before setting alarm limits and/or scaling values for that channel.
Also, the FUNC command clears all values in the Review array. FUNC
<channel>,<OFF>

FUNC <channel>,<VAC VDC FREQ>,<range>
FUNC <channel>,<OHMS>,<range>,<terminals>
FUNC <channel>,<TEMP>,<thermocouple type>
FUNC <channel>,<TEMP>,<PT>,<terminals>

<channel> = (0 .. 20)

If the channel number is invalid, an Execution Error is generated.

<function> = VAC VDC OHMS FREQ TEMP OFF

If the function is OFF, any additional supplied parameters generate a
Command Error.

<range> = AUTO (1 .. 6) (Thermocouple/RTD_type)

Ranges for VAC, VDC, OHMS, and FREQ are listed in the following
table. Enter AUTO to select autoranging. Use of any other values causes
an Execution Error.

RANGE

VOLTAGE

OHMS

FREQUENCY

1

300 mV

300

900 Hz

2

3 V

3k

9 kHz

3

30 V

30 k

90 kHz

4

150/300 V

300 k

900 kHz

5

Error

3 M

1 Mhz

6

Error

10 M

Error