Temperature Measurements - Accuracy (Thermocouples) (ITS-90)

Thermocouple

Accuracy (±°C)*

18°C to 28°C

0°C to 60°C

Type

(°C)

Temperatur

e

(°C)

90 Days

Slow

1 Year

Slow

1 Year

Fast

1 Year

Slow

1 Year

Fast

-100 to -30

0.44

0.45

0.92

0.57

1.10

J

-30 to 150

0.41

0.43

0.83

0.61

1.06

150 to 760

0.48

0.53

0.98

0.92

1.42

-100 to -25

0.53

0.54

1.14

0.66

1.33

K

-25 to 120

0.46

0.48

0.96

0.66

1.19

120 to 1000

0.79

0.85

1.55

1.45

2.20

1000 to 1372

1.32

1.42

2.30

2.29

3.23

-100 to -25

0.63

0.64

1.44

0.75

1.61

N

-25 to 120

0.53

0.54

1.21

0.66

1.39

120 to 410

0.45

0.48

1.05

0.73

1.30

410 to 1300

0.88

0.95

1.70

1.63

2.41

-100 to -25

0.46

0.47

0.92

0.59

1.10

E

-25 to 350

0.39

0.42

0.77

0.65

1.02

350 to 650

0.47

0.51

0.90

0.88

1.30

650 to 1000

0.63

0.70

1.18

1.24

1.77

-150 to 0

0.68

0.69

1.49

0.82

1.71

T

0 to 120

0.45

0.46

0.95

0.59

1.13

120 to 400

0.37

0.39

0.80

0.61

1.02

250 to 400

0.86

0.88

2.49

1.03

2.68

R

400 to 1000

0.90

0.94

2.28

1.34

2.63

1000 to 1767

1.49

1.58

3.12

2.39

3.95

250 to 1000

1.00

1.05

2.68

1.47

2.93

S

1000 to 1400

1.22

1.29

2.74

1.86

3.33

1400 to 1767

1.66

1.76

3.54

2.61

4.43

600 to 1200

1.18

1.19

3.61

1.35

3.77

B

1200 to 1550

1.14

1.19

2.57

1.60

2.99

1550 to 1820

1.33

1.39

2.86

1.93

3.41

0 to 150

0.74

0.76

1.93

0.89

2.11

150 to 650

0.69

0.73

1.66

1.08

2.04

C

650 to 1000

0.89

0.95

2.00

1.48

2.57

1000 to 1800

1.72

1.85

3.46

2.99

4.65

1800 to 2316

2.60

2.82

5.35

4.80

7.40

* Sensor inaccuracies are not included.