Fluke 2625A User Manual

Page 174

Advertising
background image

2620A/2625A

Users Manual

C-4

There are no device-to-device messages.

11.

The size of any block data responses, Section 8.7.9.4.

There are no block data responses.

12.

A list of common commands and queries which are implemented, Section 10.

*CLS, *ESE, *ESE?, *ESR?, *IDN?, *OPC, *OPC?, *RST, *SRE, *SRE?, *STB?,
*TRG, *TST?, *WAI

13.

A description of the state of the device after successful completion of the Calibration
query, Section 10.2.

The *CAL? command not implemented (an optional command).

14.

The maximum length of the block used to define the trigger macro, if *DDT is
implemented, Section 10.4.

*DDT is not implemented.

15.

The maximum length of macro labels, the maximum length of the block used to
define a macro, and how recursion is handled during macro expansion, if the macro
commands are implemented, Section 10.7.

Macros are not implemented.

16.

A description of the response to the identification common query, *IDN?, Section
10.14.

The *IDN? query returns:

FLUKE,2620A,0,M2.41 A3.7 D1.3

The version number of the main software is "M2.41", "A3.7" is the version number of
the analog sub-system software, and "D1.3" is the version number of the display sub-
system software.

17.

The size of the protected user data storage area, *PUD, Section 10.27.

*PUD not implemented. There is no protected user data storage area.

18.

The size of the resource description, if the *RDT command or *RDT? query are
implemented, Sections 10.30 and 10.31.

The *RDT and *RDT? commands are not implemented.

19.

A description of the states affected by *RST (Section 10.32), *LRN? (Section
10.17), *RCL (Section 10.29), and *SAV (Section 10.33).

*RST restores the device to the state assumed at initial power-up, except for those
items specifically forbidden by the *RST command definition. The initial power-up
state is defined under item "4.", above.

*LRN?, *RCL, and *SAV are not implemented.

20.

A description of the scope of the self-test performed by the *TST? query, Section
10.38.

* TST? performs the tests listed under "*TST?" in Table 4-8 of the User’s Manual. The

device reverts to the power-up state after performing these tests.

21.

A description of additional status data structures used in the device’s status reporting,
Section 11.

The Instrument Event Enable (IEE) register and the Instrument Event Register (IER)
are described in Figure 4-4.

Advertising
This manual is related to the following products: