Fluke 2625A User Manual

Page 114

Advertising
background image

2620A, 2625A

Users Manual

4-44

Table 4-8. Command and Query Reference (cont)

Use the lowest two bits of the value given as individual flags to specify
the temperature configuration. Therefore, the value given must be in the
range from 0 through 3 or an Execution Error is generated. These
settings affect every channel; they cannot be set for each channel
individually. If this command is attempted when measurements are
active, an Execution Error results.

Bit Number

Value

Meaning

0

0

°

C

1

°

F

1

0

Disable open thermocouple detection

1

Enable open thermocouple detection

TEMP_CONFIG?

Temperature Configuration Query

Response to this query corresponds to settings made with the
TEMP_CONFIG command, as follows:

Bit Number

Value

Meaning

0

0

°

C

1

°

F

1

0

Disable open thermocouple detection

1

Enable open thermocouple detection

TIME

Set the instrument clock.

TIME <hours>,<minutes>

<hours> = (0 .. 23) (24-hour scale, 18:00 = 6:00 pm).
<minutes> = (0 .. 59)

Invalid values generate an Execution Error. Seconds are automatically
set to 00 with the TIME command.

TIME_DATE?

Retrieve Time and Date

Returns comma-separated integer values for time, date, and year as
maintained by the instrument clock. These returned values use the TIME
and DATE command integer formats as follows:

hours

0-23

minutes

0-59

seconds

0-59

month

1-12

day

1-31

year

00-99

Seconds are returned, but cannot be set other than to 00.