Trigger input – Fluke 2625A User Manual

Page 165

Advertising
background image

Appendices

Specifications

A

A-13

Isolation

none
dc-coupled

Threshold

1.4V

Hysteresis

500 mV

Trigger Input

Input Voltages

contact closure and TTL compatible
“high” =2.0V min, 7.0V max
“low” = -0.6V min, 0.8V max

Isolation

None
dc-coupled

Minimum Pulse Width

5 µs

Maximum Frequency

5 Hz

Specified Conditions

The instrument must be in the quiescent state, with no interval scans in process, no
commands in the queue, no RS-232 or IEEE interface activity, and no front panel
activity if the latency and repeatability performance is to be achieved. For addition
information, refer to Chapter 5.

Maximum Latency

Latency is measured from the edge of the trigger input to the start of the first channel
measurement for the Specified Conditions (above).

480 ms for fast rate, scanning DCV, ACV, ohms, and frequency only
550 ms for fast rate, scanning any thermocouple or 100 mV dc channels
440 ms for slow rate, scanning DCV, ACV, ohms, and frequency only
890 ms for slow rate, scanning any thermocouple or 100 mV dc channels

Repeatability

3 ms for the Specified Conditions (above)