A. specifications, Introduction, Response times – Fluke 2625A User Manual

Page 153: Specifications, Appendix a

Advertising
background image

A-1

Appendix A

Specifications

Introduction

The instrument specifications presented here are applicable within the conditions listed
in the Environmental chapter.

The specifications state total instrument accuracy following calibration, including:

A/D errors

Linearization conformity

Initial calibration errors

Isothermality errors

Relay thermal emf’s

Reference junction conformity

Temperature coefficients

Humidity errors

Sensor inaccuracies are not included in the accuracy figures.

Accuracies at Ambient Temperatures Other Than Specified

To determine typical accuracies at temperatures intermediate to those listed in the
specification tables, linearly interpolate between the applicable 0ºC to 60ºC and 18ºC to
28ºC accuracy specifications.

Response Times

Refer to Typical Scanning Rate and Maximum Autoranging Time later in this Appendix.