Dc voltage inputs, Input impedance, Normal mode rejection – Fluke 2625A User Manual

Page 154: Common mode rejection, Maximum input, Cross-talk rejection

Advertising
background image

2620A/2625A

Users Manual

A-2

DC Voltage Inputs

Range

Resolution

Slow

Fast

300 mV

10

µ

V

0.1 mV

3 V

0.1 mV

1 mV

30 V

1 mV

10 mV

300 V

10 mV

0.1 V

Accuracy ± (% ± V)

Range

18°C to 28°C

0°C to 60°C

90 Days, Slow

1 Year, Slow

1 Year, Fast

1 Year, Slow

1 Year, Fast

300 mV

0.018% + 20 µV

0.023% + 20 µV

0.040% + 0.2 mV

0.067% + 20 µV

0.084% + 0.2 mV

3V

0.019% + 0.2 mV

0.024% + 0.2 mV

0.041% + 2 mV

0.065% + 0.2 mV

0.082% + 2 mV

30V

0.019% + 2 mV

0.024% + 2 mV

0.041% + 20 mV

0.086% + 2 mV

0.103% + 20 mV

150/300V

0.019% + 20 mV

0.024% + 20 mV

0.041% + 0.2V

0.087% + 20 mV

0.104% + 0.2V

Input Impedance

100 M

e minimum in parallel with 150 pF maximum for all ranges 3V and below 10 Me

in parallel with 100 pF maximum for the 30V and 300V ranges

Normal Mode Rejection

53 dB minimum at 60 Hz ±0.1%, slow rate
47 dB minimum at 50 Hz ±0.1%, slow rate

Common Mode Rejection

120 dB minimum at dc, 1 k

e imbalance, slow rate

120 dB minimum at 50 or 60 Hz ±0.1%, 1 k

e imbalance, slow rate

Maximum Input

300V dc or ac rms on any range for channels 0,1, and 11
150V dc or ac rms for channels 2 to 10 and 12 to 20

Cross-Talk Rejection

Refer to Appendix D