Ohms inputs – Fluke 2625A User Manual

Page 161

Advertising
background image

Appendices

Specifications

A

A-9

Ohms Inputs

Range Resolution

Typical Full

Maximum

Current

Through

Maximum Open

Slow Fast

Scale Voltage Unknown Circuit Voltage

300

10 m

0.1

0.22 V 1 mA 3.2V

3 k

0.1

1

0.25V 110

µ

A 1.5 V

30 k

1

10

0.29 V 13

µ

A 1.5 V

300 k

10

100

0.68 V 3.2 .

µ

A 3.2 V

3 M

100

1 k

2.25 V 3.2

µ

A 3.2 V

10 M

1 k

10 k

2.72 V 3.2

µ

A 3.2 V

4-Wire Accuracy ±(%±

)

Range 18°C to 28°C 0°C to 60°C

90 Days, Slow 1 Year, Slow 1 Year, Fast 1 Year, Slow 1 Year, Fast

300

0.013% + 20 m

0.014% + 20 m

0.014% + 0.2

0.031% + 20 m

0.031% + 0.2

3 k

0.015% + 0.2

0.016% + 0.2

0.016% + 2

0.039% + 0.2

0.039% + 2

30 k

0.013% + 2

0.014% + 2

0.014% + 20

0.039% + 2

0.039% + 20

300 k

0.020% + 20

0.021% + 20

0.021% + 200

0.050% + 20

0.050% + 200

3 M

0.059% + 200

0.063% + 200

0.063% + 2 k

0.231% + 200

0.231% + 2 k

10 M

0.168% + 2 k

0.169% + 2 k

0.709% + 20 k

0.573% + 2 k

0.923% + 20 k

2-Wire Accuracy

Add 4.0

to accuracy specifications for channels 1 to 20, and add 20 m

for channel 0.

Lead wire resistances are not included.

Input Protection