Typical scanning rate – Fluke 2625A User Manual

Page 163

Advertising
background image

Appendices

Specifications

A

A-11

Typical Scanning Rate

See table below. The measurement conditions are: averaged rate over 20 scans;
continuous scanning; alarm limits and Mx+B scanning set on all channels; logging data
to internal memory; and RS-232 communications set at 9600 baud. Measurements were
taken with short-circuit inputs on all channels, except frequency, which was taken with
5V at 15 Hz on all channels.

CHANNELS

FUNCTION RANGE SLOW FAST

1 10 20 1 10 20

VDC 300 mV 1.8 3.9 4.1 2.5 13.2 18.3

3V 1.8 3.9 4.1 2.5 13.2 18.4

30V 1.8 3.9 4.1 2.5 13.2 18.2

150/300V 1.8 3.9 4.1 2.5 13.2 18.1

AUTO 1.8 3.6 3.9 2.4 11.3 14.1

TEMPERATURE J (TC) 1.0 3.3 3.8 2.1 10.9 15.2

PT (RTD) 1.7 3.1 3.2 2.1 6.0 6.7

VAC 300 mV 1.1 1.5 1.6 1.3 2.5 2.6

3V 1.1 1.5 1.6 1.3 2.5 2.6

30V 1.1 1.5 1.6 1.3 2.5 2.6

150/300V 1.1 1.5 1.6 1.3 2.5 2.6

AUTO 1.0 1.5 1.5 1.3 2.4 2.5

OHMS 300

1.7 3.1 3.2 2.1 6.0 6.7

3 k

1.7 3.1 3.2 2.1 6.0 6.7

30 k

1.7 3.1 3.2 2.1 6.0 6.7

300 k

1.2 1.8 1.8 1.8 4.1 4.4

3 M

1.1 1.6 1.7 1.8 3.9 4.2

10 M

1.1 1.6 1.6 1.7 3.8 4.0

AUTO 1.7 3.1 3.2 2.1 6.0 6.7

FREQUENCY ANY 0.5 0.6 0.6 0.6 0.7 0.7