Digital and Alarm Outputs

Output Logic Levels

Logical “zero”:

0.8V max for an Iout of -1.0 mA (1LSTTL load)

Logical “one”:

3.8V min for an Iout of 0.05 mA (1LSTTL load)

For non-TTL loads: 1.8V max for an Iout or -20 mA
Logical “zero”:

3.25 max for an Iout of -50 mA

Isolation

none

Real-Time Clock and Calendar

Accuracy

Within 1 minute per month for 0°C to 50°C range

Battery Life

10 years minimum for Operating Temperature range

Environmental

Warmup Time

1 hour to rated specifications

15 minutes when relative humidity is kept below the rated maximum minus 20% (e.g.
below 70% for a 90% maximum rating).

Operating Temperature

0°C to 60°C

Storage Temperature

-40°C to +75°C
Instrument storage at temperature extremes may necessitate adding up to
0.008% to the dc and ac voltage accuracy specifications. Alternatively, any resulting
shift can be compensated for by recalibrating the instrument.

Relative Humidity (Non-Condensing)

90% maximum for 0°C to 28°C
75% maximum for 28°C to 35°C
50% maximum for 35°C to 50°C
35% maximum for 50°C to 60°C

(Except 70% maximum for 0°C to 35°C, 30% maximum for 40°C to 50°C, and 20%
maximum for 50°C to 60°C, for the 300 k

e, 3 Me, and 10 Me ranges.)