Mx+b scaling, Setting the scan interval – Fluke 2625A User Manual

Page 39

Advertising
background image

Overview

Setting the Scan Interval

2

2-15

Mx+B Scaling

The menu for setting Mx+B scaling values takes you through the following steps:

Set the M value (sign and number).

Set the decimal point location for the M value.

Set the multiplier for the M value (m, x1, k, M).

Set the B value (sign and number).

Set the decimal point location for the B value.

Set the multiplier for the B value (m, x1, k, M).

Use

G, D, H and J to cycle through the selections at each step. Then press

EE to accept your choice and advance to the next step.
For example, to set an M value of 1.8 and a B value of 32 for channel 7, do the
following:

1.

Select channel 7 by pressing

G or D.

2.

Press

B to begin Mx+B scaling setting for channel 7.

3.

Use the arrow buttons to select "+18000". Press

H or J to move between digits.

Press

G or D to select the value for each digit. (Ignore the decimal point). Press

EE to accept these digits.

4.

Use the arrow buttons to select the decimal point position ("+1.8000"). Then press

E to accept this position.

5.

Next, use the arrow buttons again to select the multiplier ("x1"). Press

E to

accept this multiplier.

6.

Use the arrow buttons to select +00320. Then press

EE.

7.

Use the arrow buttons to select the decimal point position ("+0032.0"). Then press

E to accept this position.

8.

Next, use the arrow buttons again to select the multiplier ("x1").Press

E to

accept this multiplier as the multiplier for the B value.

Mx+B scaling is now set at 1.8x+32 for channel 7.

Pressing

D while partway through setting the M value undoes all changes entered thus

far, returning the instrument to Inactive Mode. Pressing

D while partway through

setting the B value cancels entries made thus far for the B value, but does not affect any
changes already made to the M value.

Setting the Scan Interval

Press

I to set the scan interval. The latest value appears on the right display. The

format of the display is H:MM:SS. Use

J and H to move between digits. Use G and

D to select the new value for a digit. Values ranging from 0:00:00 to 9:99:99 are
allowed. Press

E to accept the displayed value.

Set the interval to 0:00:00 for continuous scanning. If you set the interval to a value
shorter than the time required to measure all channels, scanning effectively becomes
continuous.