Selecting scan interval, Selecting the measurement rate – Fluke 2625A User Manual

Page 57

Advertising
background image

Operating the Instrument from the Front Panel

Instrument Configuration

3

3-15

This number editing technique occurs during Configuration Mode operations whenever
you are setting a numeric value. Instances of number editing include the following:

R0 (RTD)Table 3-6

Alarms Table 3-7

Mx+B Scaling Table 3-9

Scan Interval TimeTable 3-10

Time/Date Table 3-13

Selecting Scan Interval

The scan interval is the period between starts of measurement scans. The resolution of
the scan interval is one second; the maximum scan interval is 9 hours: 99 minutes: 99
seconds. The format of the scan interval is x:xx:xx; 1:30:45 translates to 1 hour, 30
minutes and 45 seconds.

If scan interval of "0:00:00" is specified, continuous scanning will occur when the Scan
function is activated. If more than a few seconds are required to scan all defined
channels, a short scan interval (a few seconds) effectively becomes continuous scanning.

If the Monitor function is turned on when the instrument is continuously scanning, a
Monitor channel measurement is still taken between each scan.

Table 3-10 illustrates the button sequence used for setting scan interval. If necessary,
refer to "Entering and Changing Numeric Values" for a more detailed description of the
number editing technique used here.

Table 3-10. Scan Interval

Hour

Minute

Second

PRESS
THESE
BUTTONS:

I

G

D

J

G

D

J

G

D

J

G

D

J

G

D

E

TO
CHANGE
THE
INDICATED
DIGIT:

0

:00:00

0:

0

0:00

0:0

0

:0

0

0:00:

0

0

0:00:0

0

Note: Go backward (from SECOND to MINUTE to HOUR) by pressing

H

.

Press

E

at any time to accept the displayed scan interval and exit scan interval selection. Press

C

at any time to exit interval selection without storing any changes.

Selecting the Measurement Rate

The slow measurement rate provides the highest accuracy and resolution. A fast rate can
be selected, but keep in mind that fast rate provides one less digit of resolution than does
slow rate. Measurement rate selection is illustrated in Table 3-11.

Advertising
This manual is related to the following products: