Memory storage: computer interface control, Memory retrieval – Fluke 2625A User Manual

Page 76

Advertising
background image

2620A, 2625A

Users Manual

4-6

10:33:45 5/11/90

Lines following the time and date contain measurement data for channels that have been
set up for this session. The last line of the printout contains the Totalizer count and the
status of the digital I/O lines.

Channel data is formatted to fit three readings onto an 80-column line. Each reading
provides the following information:

Channel number, followed by a colon and a space (4 characters)

space or minus sign (1 character)

Digits and decimal point. Measurements taken at the fast rate use 5 characters (4
digits plus a decimal point.) Measurements taken at the slow rate use 6 characters (5
digits plus a decimal point.)

Space(1 character)

Measurement multiplier - ’m for X0.001, space for X1, ’k’ for X1,000, ’M’ for
X1,000,000

Measurement units - VDC, VAC, OHMS, HZ, C, F, Mx+B (4 characters, left
justified)

Space(1 character)

Alarm indication for Limit-1 (H=High, L=Low, R=Return) (1 character); a blank
space = no alarm limit is used, or alarm inactive.

Slash (/) used as a separator when either alarm indicator is shown (1 character)

Alarm indication for Limit-2 (H, L, R); space = no alarm limit is used, or alarm
inactive.

Two spaces between channel measurements (2 characters)

The following example illustrates this format:

07:42:0102/09/91

5: -3.2345 mMX+B H/R12:123.87 kOHMS/L17:31.268VAC

ALM:15DIO:204TOTAL:0

Memory Storage: Computer Interface Control

Memory Storage can be controlled from the front panel or over the RS-232 interface.
Refer to Chapter 3 for front panel operation.

Through the RS-232 interface, the PRINT_TYPE and PRINT commands can be used to
set up and enable Memory Storage. PRINT_TYPE selects the destination (0 for printer, 1
for Memory Storage, and 2 for both) and the mode used to select the scans to store (0 for
ALL, 1 for ALAr, or 2 for trAnS.) PRINT sets storage on (1) or off (0).

The PRINT_TYPE? and PRINT? queries can be used to determine present settings for
the PRINT_TYPE (destination, type) and PRINT commands, respectively.

Memory Retrieval

Note

The LOG? query does not return channel numbers with scan data.
Therefore, if you add or delete defined channels, you may want to clear the
memory contents so that subsequent LOG? queries return only data for the
new set of scanned channels.

Advertising
This manual is related to the following products: