Step 4 — o, Perate – Millennium Enterprises Net Safety MLP-A/AR User Manual

Page 15

Advertising
background image

Net Safety Monitoring Inc.

MLP-A/AR/AD-ST1400-25

10

STEP 4 — O

PERATE

The Controller faceplate contains most functional elements of the user interface. Below is a description of that functionality.

Figure 12: Controller Functionality

Status

Red

Slow Flash:
Gas present
Flast flash :
Cal or Fault
Solid:
Overrange

Pull

Here

Green

Short Blips:
Normal
operation.
No gas
present

Cal/

Res

et

Ma

gne

t

On

Sid

e

Off

Power

On

Pull
Here

Current O/P Check

Connect current probes

to meter jacks and

read mA output

Calibration/Setup

-Hold Cal/Reset or
magnet until countdown
is zero (~10 sec)
-Follow instruction on
display

Magnetic Reed Switch - provides non-intrusive access for
programming, calibrating and resetting.

-Place magnet against Housing (where indicated) for less than one
second to initiate a basic system reset (clear a latched alarm) and
make selections.

-Place magnet against Housing (where
indicated) and hold to program, view
current settings and calibrate.

Refer to "Magnetic Reed Switch" on page 12
for further information.

Scrolling 8-character display -
provides various status messages and
prompts. Refer to "Status LEDs,
Display Messages and Current
Loop" on page 11.

Pull Here knob - unscrew the two
screws and gently pull to remove
faceplate. Removal allows access
to terminal boards. The faceplate
remains attached by the ribbon
cable.

Status Light (Red and Green) -
indicates status of unit. Refer to
"Status LEDs, Display Messages and
Current Loop" on page 11 for
detailed explanation of states/status.

Current Test Jacks - to take current loop measurements use the following procedure:

-Insert your current meter leads into the Test Jacks.

-Set external devices to bypass, if necessary, to avoid unwanted alarm response

-Review current loop measurements.

Refer to "Current Loop Measurement (Test Jacks)" on page 12 and Table 2, "Status LEDs, Display
Messages and Current Loop", on page 11.

Calibration Button - provides intrusive
access for programming, calibrating and
resetting.

-Push for less than one second to initiate a
basic system reset (clear a latched gas
alarm) and make selections.

-Push and hold to program, view current
settings and calibrate.

Refer to "Calibration Button" on page 11
for further information.

ON/OFF Switch - used to turn Controller on and
off. Housing must be removed to access.

Pull Here knob - unscrew the two screws and
pull to remove faceplate. Removal allows
access to PCBs. The faceplate remains
attached by the ribbon cable.