STEP 5 — C

ALIBRATE

C

ALIBRATION

P

ROCEDURE

WARNING:

The calibration procedure requires about 5 minutes to complete. If gas is not applied at the appropriate time, a calibration failure may occur (refer

to "Calibration Failure" on page 14 for specific information).

For accurate performance, the concentration of gas, corresponding to 0 to 25% oxygen is converted to a linear 4 to 20 mA output signal which can be powered from
the primary dc supply of the instrument. Power up the unit for at least 4 hours BEFORE first calibration.

The following calibration procedure should be followed to ensure an accurate correlation between the 4 to 20 mA output signal and the gas concentration.

Calibration Procedure

1. Confirm successful power up of Controller

LED Green Blip/blink every 2 seconds; no fault indicated.

2. Flow certified ZERO AIR at a rate of 0.5 litres per minute through the calibration cup for 1 minute to ensure clean air environment.
3. Use the Calibration Button or Reed Switch to access the Main Menu and wait for countdown to complete.
4. When

Calibrate Sensor YES?

displays, use the Calibration Button or Reed Switch to select

YES?

Selection will be confirmed by a flashing

YES

.

5. When

Apply Clean Air

displays apply clean air.

6. When

Apply 20.9% O2

displays, apply O2 at a rate of 0.5 litres per minute using the calibration cup.

7. The display will show

Calibrating

as the internal settings are being modified (current loop 3.3 mA).

8. Remove span gas when the message

Remove Gas

displays (Status LED Green solid and 3.6 mA output).

9. The message

Cal Complete

will display when calibration is complete.

10. Apply zero gas (clean air) again to purge system.

Note:

Always apply test gas after calibration to verify operation.

Figure 13: Calibration Procedure

Select

YES?

Display
Shows

Action

Calibrate

Sensor YES?

Apply clean
(zero air)

Apply Clean

Air

Apply 20.9%

O2

Apply O2

Calibrating

Remove Gas

Remove gas

Cal Complete