Net Safety Monitoring Inc

MLP-A/AR/AD-ST1400-25

16

M

ODBUS

O

PTIONS

The digital RS-485 Modbus RTU protocol is used. There are two Modbus
options: Node Address and Baud Rate.

Node Address

Each device connected to the chain must be assigned a unique node address (1-
255). The last number in the address is selected first.

Step 1:

When

Modbus Setup? YES?

displays press the Calibration Button

or use the Reed Switch to select.
The flashing

YES

confirms the selection.

Step 2:

The current node address will display

Node: 000

.

Step 3:

Wait for the prompt

New Address? YES?

and press the Calibration

Button or use the Reed Switch to select.

Step 4:

Use the Calibration Button or Reed Switch to select each of three
numbers in the new address:
- select the last number in the address first:

0

thru

9

.

- select the next number in the address:

0

thru

9

.

- select the first number in the address last:

0 1 2

.

Baud Rate

The transmission speed must be defined.

Step 1:

When

Modbus Setup? YES?

displays press the Calibration Button

or use the Reed Switch to select.
The flashing

YES

confirms the selection.

Step 2:

After setting the Node Address, the current Baud Rate will display

XX.X BPS

.

Step 3:

Wait for the prompt

New Baud Rate? YES?

and press the

Calibration Button or use the Reed Switch to select.

Step 4:

The available baud rates will display:

2400s

,

4800s

,

9600s,14.4s

,

19.2s

,

28.8s

,

38.4s

,

57.6s

.

Step 5:

Use the Calibration Button or use the Reed Switch to select baud
rate when it displays.

Step 6:

The flashing

YES

confirms the selection.

E

NTER

R

ESTRICTED

M

ENU

The Restricted Menu allows you to set the Low Power Options.

Step 1:

Press and hold the Calibration Button or the magnet to the Reed
Switch to enter the Main Menu; wait for the countdown, from 10 to
0, to end.

Step 2:

When

Enter restricted menu YES?

displays press the

Calibration Button or use the Reed Switch to select.
The flashing

YES

confirms the selection.

Step 3:

When

Are you sure? YES?

displays press the Calibration Button

or use the Reed Switch to select.
The flashing

YES

confirms the selection.

Set Low Power Options - Optional

Two optional low power features are available to reduce overall power
consumption. Use this option to:

• Dim the LED display although display will still be visible in most lighting

conditions (default setting)

• Disable the 4 to 20 mA analog output for applications requiring only relay

output (default setting)

Step 1:

Enter the Restricted Menu (refer to "Modbus Options" on page 16).

Step 2:

When

Set Low Power Options YES?

displays press the Calibration

Button or use the Reed Switch to select.
The flashing

YES

confirms the selection.

Step 3:

When

Lower Display Brightness YES?

displays press the

Calibration Button or use the Reed Switch to select or wait until

Display Full Brightness YES?

displays and select.

The flashing

YES

confirms the selection.

Step 4:

When

Disable 4 - 20mA O/P YES?

displays press the Calibration

Button or use the Reed Switch to select or wait until

Enable 4 -

20mA O/P YES?

displays and select.

The flashing

YES

confirms the selection.