Calibration Failure

If the calibration procedure fails, the display shows

Fail Cal

, the Status LED

alternates Red and Green flashes and the analog output changes back and forth
from 3.0 to 3.3 mA.

The unit remains in a failed state until manually reset. After the Manual Reset,
the unit will return to normal operation based on previous calibration values.
Refer to "Manual Reset" on page 17 for instructions.

STEP 6 — M

ONITOR

R

EVIEW

R

ELAY

S

ETTINGS

This is a read-only mode; changes cannot be made.

Step 1:

Press and hold the Calibration Button or the magnet to the Reed
Switch to enter the Main Menu; wait for the countdown, from 10 to
0, to end.

Step 2:

When

Review Relay Settings

displays press the Calibration

Button or use the Reed Switch to select.
The flashing

YES

confirms the selection.

Step 3:

The Fault Alarm is fixed (

Energized

/

Non-Latching

) and

displays first. Then the Low Alarm level, coil and latch status
display, followed by the High Alarm level, coil and latch status.

Step 4:

At this point, the option to

Set Relay Options YES?

is

displayed.