Air purging – Sanyo CHX03652 User Manual

Page 191

Advertising
background image

5-3

Test Run and Others

1

2

3

4

5

6

7

8

(3) Do a leak test of all joints of the tubing (both

indoor and outdoor) and both gas tube and liquid

tube service valves. Bubbles indicate a leak. Wipe

off the soap with a clean cloth after the leak test.

(4) After the system is found to be free of leaks,

relieve the nitrogen pressure by loosening the

charge hose connector at the nitrogen cylinder.

When the system pressure is reduced to normal,

disconnect the hose from the cylinder.

Evacuation

(1) Attach the charge hose end described in the

preceding steps to the vacuum pump to evacuate

the tubing and indoor unit. Confirm that the “Lo”

knob of the manifold valve is open. Then, run the

vacuum pump. The operation time for evacuation

varies with the tubing length and capacity of the

pump. The following table shows the amount of

time for evacuation:

The required time in the above table is calculated

based on the assumption that the ideal (or target)

vacuum condition is less than -14.7 psig (–755 mm

Hg, 5 Torr).

(2) When the desired vacuum is reached, close the

“Lo” knob of the manifold valve and turn off the

vacuum pump. Confirm that the gauge pressure is

under -14.7 psig (–755 mmHg, 5 Torr) after 4 to 5

minutes of vacuum pump operation.

Manifold valve

Pressure
gauge

Lo

Hi

Vacuum pump

Outdoor unit

Liquid
tube

Gas
tube

Close

Open

Close

Open

Service port ø5/16''

Required time for evacuation

when 30 gal/h vacuum pump is used

If tubing length is

If tubing length is

less than 49 ft.

longer than 49 ft.

45 min. or more

90 min. or more

NOTE

Fig. 5-4

1. Air Purging

Advertising
This manual is related to the following products: