TROUBLESHOOTING & REPAIR

F-13

F-13

V205-T AC/DC

INPUT FILTER CAPACITOR DISCHARGE PROCEDURE

Service and repair should be performed by only Lincoln Electric factory trained
personnel. Unauthorized repairs performed on this equipment may result in danger to
the technician or machine operator and will invalidate your factory warranty. For your
safety and to avoid electrical shock, please observe all safety notes and precautions
detailed throughout this manual.

If for any reason you do not understand the test procedures or are unable to perform the
test/repairs safely, contact the Lincoln Electric Service Department for electrical
troubleshooting assistance before you proceed. Call 1-800-833-9353 (WELD).

DESCRIPTION

This procedure will drain off any charge stored in the four capacitors that are part of the
main IGBT inverter board assembly. This procedure MUST be performed, as a safety
precaution, before conducting any test or repair procedure that requires you to touch
internal components of the machine.

MATERIALS NEEDED

Phillips Screwdriver

Insulated Pliers

Insulated Gloves

High Wattage Resistor (25-1000 ohms and 25 watts minimum)

Volt Meter