Operator Manual, Supercharger 60

R3.0

Page 14 of 62

3.1.12.1. 3 AMPS idling plus ½ the value of the charging current; e.g. when

charging 40 AMPS the line current is 23 AMPS (Note: nominal values,
line voltage and frequency dependent).

3.2 DISCHARGE

3.2.1. CONTROL: Transistors, air cooled.

3.2.2. MAXIMUM CURRENT:

3.2.2.1. 60 AMPS. Total power dissipation must not exceed 1.2KW (1200

watts, the product of battery voltage and battery current).

3.2.2.2. NOTE: Discharge current is automatically limited to 30 AMPS for

battery voltages in excess of 32 VOLTS. A current fault will be
generated if attempting to discharge at more than 30 AMPS.

3.2.3. MINIMUM INCREMENT:

3.2.3.1. 0.1 AMP.

3.2.4. ACCURACY and STABILITY of settings:

3.2.4.1. ±1% of reading, ±0.1AMP.

3.2.5. MODES:

3.2.5.1. Analysis (automatic cut-off)

3.2.5.2. Deep cycle (full discharge).