NOTE 3: Insert an external shunt/ammeter in series with any of the

battery leads (Use the Single Cell adaptor to facilitate the
connection or simply use the CAL-100 which is a calibrator
designed specifically to test and certify Charger-Analyzer Battery
Charger-Analyzers.

NOTE 4: The charge output current wave is in the form of short

pulses (at twice the line frequency). The use of clamp-on meters
and moving magnet meters will yield erroneous readings. Verify in
discharge (pure DC) if better meters are not available.

9.2.2. VOLTMETER.

9.2.2.1. ZERO: No zero adjustments.

9.2.2.2. SCALE:

• This meter has a full scale of 1.999 volts, but an attenuator on the

board converts it to a full scale of 1.999, 19.99 or 199.9 volts, as
selected by the DPM selector switch.

• Verify and calibrate using a voltage source (Power Supply or

CAL-100, as a battery voltage simulator) between 10 and 20 volts
in the "EXTERNAL 20V" position.

• Adjust R17 for a reading within " 0.01V of the reference voltmeter.
NOTE: individual scales cannot be adjusted.