2.1.3 CHANNEL CONNECTORS

The five input connectors on the right side of the OM-550 accept either voltage or
thermocouple inputs values.

Figure 2-9

The mating connectors are color coded by conductor material, as described in the table
below. For best accuracy, use the appropriate color of connector.

CONNECTOR COLOR

APPLICATION

White

Uncompensated

Voltage inputs

Black

Type J

thermocouples

Yellow

Type K

thermocouples

Blue

Type T

thermocouples

Purple

Type E

thermocouples

Orange

Type N

thermocouples

••

Optional Connectors:

White

Type B

thermocouples

Green

Type S or R

thermocouples

To minimize danger to personnel, and to avoid ground loops which could affect
measurement accuracy, always disconnect the RS-232 cable before attaching the
input signals.