2.3 INPUTS

As previously discussed, the OM-550 accepts either voltage or thermocouple inputs.
The Signals discussed in this section are input through one of the channel connectors
on the right side of the OM-550.

2.3.1 VOLTAGE (TRANSDUCER) INPUTS

The Voltage connector used for inputs is a white connector with copper pins.

Figure 2-13

2.3.2 PROBES

Optional Voltage and Current probe sets are available for the OM-550. They provide a
convenient way to connect the OM-550 to voltage or current signals:

PROBE SET

APPLICATION

±2 VDC

VOLTAGE

connects to signals in the ±2 VDC range

±10 VDC VOLTAGE

connects to signals in the ±10 VDC range

4-20 mA

CURRENT

converts 4-20 mA loop currents to ±2 VDC signals

For convenience, the OM-550 can be programmed to Scale data. The following table
includes m (slope), b (offset) and dp (decimal point) settings for common probe
applications:

PROBE INPUT RANGE m b dp PRINT RANGE UNITS

±2 V

±2 VDC

+1.00000

+000000

4

-2.0000 to +2.0000

V*

±10 V

±10 VDC

+0.50000

+000000

3

-10.000 to +10.000

V

0-10 VDC

+0.50000

+000000

3

00.000 to +10.000

V

0-5 VDC

+0.50000

+000000

3

00.000 to+05.000

V

1-5 VDC

+0.50000

+000000

3

+01.000 to +05.000

V

4-20 mA

4-20 mA

+1.00000

+000000

3

+04.000 to +20.000

mA**

* Default label for VOLTAGE channels ** Example only; many UNIT labels are available.

Refer to sections 4.15 UNIT LABELS and 4.14 VOLTAGE INPUT SCALING for more
information.