Checks and Adjustments

5- 12

1705A Spectrum Monitor

10. Check L--Band Linearity

REQUIREMENT

-- One marker per Division

±1 minor Division.

a.

Connect the output of the leveled sine wave generator, through the comb
generator, to the 1705A L--BAND INPUT. See Figure 5-4.

SG503

Comb
Generator

VARIABLE
AUTOTRANSFORMER

L--BAND
INPUT

1705A

Figure 5-4: Equipment connections for L- Band checkout

b.

Set the leveled sine wave generator Frequency to 100 MHz, Amplitude
Multiplier to X1, and the Amplitude to 1.5 (8 dBm).

c.

Set the 1705A INPUT to L--BAND.

d.

Set the SPAN/DIV to FULL.

e.

CHECK

-- for 10 frequency markers. (Ignore any sweep start marker.)

f.

CHECK

– that each marker is within one minor Division of a major

graticule line when the 2

nd

marker is on the first graticule line in from

the left. See Figure 5-5.