Checks and Adjustments

1705A Spectrum Monitor

5- 27

e.

ADJUST

-- R239 (IF AGC) for maximum signal amplitude.

f.

ADJUST

-- R240 (RF AGC) so the signal is one major Division (

±0.5

major Divisions) down from the top graticule line.

g.

ADJUST

-- R232 to its midpoint.

18. Adjust 2 dB/Div Position

a.

Disconnect the signal from the L--BAND INPUT.

b.

Turn VERTICAL POSITION fully clockwise.

c.

Hold the VIDEO push button down until 2 dB/Div appears on the crt
readout.

d. ADJUST

-- R140 to position the noise floor to center screen.

e.

Hold the VIDEO push button down again to exit the 2 dB/Div mode.

19. Adjust 590 MHz Notch

a.

Select L--BAND INPUT and set the SPAN/DIV to 10 MHz.

b.

Connect the signal generator to the L--BAND INPUT.

c.

Set the signal generator Frequency to 1200 MHz, and Amplitude to
--20 dBm.

d.

Use the HORIZONTAL POSITION control to place the 1200 MHz
signal at the extreme left graticule mark. A small spur should be visible
near the extreme right graticule mark. Note: R240, RF AGC, may be
adjusted to ease visibility of the spur. Be sure to reset R240 (step 17
part f.).

e.

ADJUST

-- C69 carefully until spur is minimized. Note: While

adjusting C69, note the 1200 MHz amplitude. The spur should be
minimized without reducing the 1200 MHz spike by more than 2 dB.

End of Adjustment Procedure