Tektronix 1705A User Manual

Page 26

Advertising
background image

Introduction

1705A Spectrum Monitor

1- 7

Table 1- 6: Certifications and Compliances (cont.)

FCC Compliance

Emissions comply with FCC Code of Federal Regulations 47, Part 15, Subpart B, Class A Limits

Installation (Overvoltage)
Category

Terminals on this product may have different installation (overvoltage) category designations. The
installation categories are:

CAT III

Distribution-level mains (usually permanently connected). Equipment at this level is
typically in a fixed industrial location.

CAT II

Local-level mains (wall sockets). Equipment at this level includes appliances, portable
tools, and similar products. Equipment is usually cord-connected.

CAT I

Secondary (signal level) or battery operated circuits of electronic equipment.

Pollution Degree

A measure of the contaminates that could occur in the environment around and within a product.
Typically the internal environment inside a product is considered to be the same as the external.
Products should be used only in the environment for which they are rated.

Pollution Degree 1

No pollution or only dry, nonconductive pollution occurs. Products in
this category are generally encapsulated, hermetically sealed, or
located in clean rooms.

Pollution Degree 2

Normally only dry, nonconductive pollution occurs. Occasionally a
temporary conductivity that is caused by condensation must be
expected. This location is a typical office/home environment.
Temporary condensation occurs only when the product is out of
service.

Pollution Degree 3

Conductive pollution, or dry, nonconductive pollution that becomes
conductive due to condensation. These are sheltered locations where
neither temperature nor humidity is controlled. The area is protected
from direct sunshine, rain, or direct wind.

Pollution Degree 4

Pollution that generates persistent conductivity through conductive
dust, rain, or snow. Typical outdoor locations.

Safety Standards

U.S. Nationally Recognized
Testing Laboratory Listing

UL1244

Standard for electrical and electronic measuring and test equipment.

Canadian Certification

CAN/CSA C22.2 No. 231

CSA safety requirements for electrical and electronic measuring and
test equipment.

European Union Compliance

Low Voltage Directive 73/23/EEC, amended by 93/69/EEC

EN 61010-1

Safety requirements for electrical equipment for measurement,
control, and laboratory use.

Additional Compliance

IEC61010-1

Safety requirements for electrical equipment for measurement,
control, and laboratory use.