Tektronix 1705A User Manual

Page 140

Advertising
background image

Maintenance

1705A Spectrum Monitor

6- 15

b.

Table 6--4 lists the signal present in a properly functioning control
circuit.

Table 6- 4: Control Circuit Test Points

Circuit Location

Signal

U5, pin 1

Approximately 5 Vdc

U5, pin 2

Approximately 2 Vdc

U5, pin 3

0 V

U5, pin 4

80 kHz triangle wave, 2 V p-p

U5, pin 6

80 kHz square wave, 18 V p-p

U3, pin 1

80 kHz square wave, 5 V p-p

U3, pin 2

2.1 Vdc

U3, pin 6

2.9 Vdc

U3, pin 7

Approximately 5 Vdc

U4, pin 13

80 kHz repetition rate, 300 ns width, approximately 3 V p-p

Q8, collector

80 kHz repetition rate, 400 ns width, switching from 5 V to
approximately 2 V.

5.

Error Amplifier Check

a.

Connect the negative output from the variable DC Power Supply to TP1.
Connect the positive output to W1 (+5 V).

b.

Connect the negative output of another variable DC Power Supply to
TP1. Connect the positive output to W4 (+15 V). Set the variable
power supply to 20 Vdc.

c.

Connect the digital multimeter between TP1 and the cathode of CR15.

d.

Set the variable DC Power Supply connected to W1 (+5 V) to 4.8 V.
The cathode of CR15 should be approximately 20 V.

e.

Set the variable DC Power Supply connected to W1 (+5 V) to 5.2 V.
The cathode of CR15 should be approximately 2 V.

f.

If this check did not reveal the cause for the +5 V supply not regulating,
refer to the Output Check and the Control Circuit Check.

Table 6--5 lists the High Volts Supply fault symptoms and procedures.

High Volts Supply