1705A Spectrum Monitor

5- 1

Section 5
Checks And Adjustments

This section consists of two separate procedures. The first, a Performance
Check, is used to determine compliance with the Performance Requirements in
the Specification. The second is the Adjustment Procedure, that provides the
instructions on how to adjust the instrument and return it to operation within the
specification.

In both procedures, front- and rear-panel controls and connectors, on the
instrument under test, are fully capitalized (e.g., 300 kHz RESOLUTION).
Control and connector names on test equipment and internal controls and
adjustments for the instrument under test are initial capitalized (e.g., Time/Div,
Geometry, etc.).

Limits, tolerances, and waveforms given in this section are guides to adjustments
and checks. They are not instrument specifications, except when listed in the
Performance Requirement column of the Specification Tables in Section 1 of this
manual.

Recommended Equipment List

The following equipment and accessory items are required to do the Performance
Check and/or Adjustment Procedures. Broad specifications are followed by a
piece of equipment that meets these specifications; in most cases, the recom-
mended instrument was used in preparing the procedures that follow.

1.

Test Oscilloscope

Vertical Amplifier:

30 MHz Bandwidth, 1 mV Sensitivity.

Time Base:

10 ns/Division to 5 ms/Division sweep speeds, triggering to 5 MHz.

For example: a TEKTRONIX 7603 Oscilloscope with a 7A18 Dual-Trace
Amplifier and a 7B53A Dual Time Base, or a TEKTRONIX 11403A
Oscilloscope with a 11A34V Video Amplifier and an 11T5H Video Trigger.
Also a 10X probe, P6106 (Tektronix Part No. 010-6106-03).