SECTION 13—BATTERIES

Part No. 1143206

59

Pronto® M41 with SureStep®

ƽ WARNING

Batteries with terminal configuration as shown below MUST be used. Batteries that

have the reverse terminal configuration MUST NOT be used - otherwise injury and

damage may occur.

NOTE: Recommended battery type is spill proof and requires no maintenance except routine 
charging.

NOTE: Charge batteries daily. DO NOT allow batteries to completely discharge. To charge 
batteries, refer to Charging Batteries on page 63.

NOTE: Invacare recommends that both batteries be replaced if one battery is defective.

Removing/Installing the Batteries

NOTE: For this procedure, refer to FIGURE 13.1 on page 60.

ƽ WARNING

Install protective caps on POSITIVE (+) and NEGATIVE (-) terminals.
All battery terminal covers (two on the right battery and two on the left battery)

MUST be installed prior to use.

Removing

1. Disconnect the joystick. Refer to Disconnecting/Connecting the Joystick on page 56.

2. Remove the seat. Refer to Removing/Installing the Seat Assembly on page 37.

3. Remove the top shroud. Refer to Removing/Installing the Top Shroud on page 49.

4. Disconnect the right battery from the controller (BLACK connector). 

DO NOT USE

NEGATIVE (-)

Battery Terminal

POSITIVE (+)

Battery Terminal

U1 Battery

U1 Battery

POSITIVE (+) Battery Terminal

NEGATIVE (-)

Battery Terminal

USE THIS CONFIGURATION