VEGA VEGAFLEX 52V User Manual

Page 10

Advertising
background image

10

VEGAFLEX 51V and 52V

Product description / Technical data

Materials

Housing

PBT (Valox) or
Aluminium (powder coated)

Cable

1.4401 (stst; 316) or
galvanized steel - PA-coated

Process connections

VEGAFLEX 51, 52

G 1

1

/

2

A, 1

1

/

2

“ NPT

steel chromated or of 1.4301 (stst)

Connection lines

- supply and signal via one two-wire cable,

the line resistance depends on the supply voltage (see diagram).

Cross-section area of conductor

generally 2,5 mm

2

Earth connection

max. 4 mm

2

Cable entry

2 x M20 x 1,5 (cable diameter 5 … 9 mm)

Voltage limit
not-Ex-sensor

Voltage limit
Ex-sensor

Supply voltage

CE-conformity

VEGAFLEX sensors meet the protective regulations of EMVG (89/336/EWG) and NSR
(73/23/EWG).
The conformity has been judged acc. to the following standards:
EMVG Emission

EN 50 081 - 1: 1992

Susceptibility

EN 50 082 - 2: 1995

NSR

EN 61 010 - 1: 1993

Display indication

Display

scalable analogue and digital measured
value indication (option).

An external measured value indication, supplied by the sensor, can be mounted up to
25 m separated from the sensor.

Advertising