VEGAFLEX 51V and 52V

Electrical connection / Connection of the sensor

3.2 Connection of the sensor

Loosen the closing screw on the upper
side of the sensor.

Open the sensor cover.

Remove the compression nut of the Pg
and shift the nut a little bit over the con-
nection cable.

Remove the rubber seal out of the Pg and
shift the seal over the connection cable.

Remove the cover of the connection cable
over a length of approx. 10 cm.

Loop the cable through the Pg into the
sensor.

Two-wire sensor

Connect the cable.
Push the white buckets of the spring
terminals with a small screwdriver and
insert the copper core of the connection
line into the terminal opening.
Check the position of the cables in the
terminal by slightly pulling.

Fasten the compression nut on the Pg.

VBUS

To the indicating instrument in the
sensor cover or to the external indicat-
ing instrument VEGADIS 50

M20 x 1,5
(diameter of the
connection cable
5…9␣ mm)