6 Setup with the indicating and

adjustment module PLICSCOM

6.1 Short description

The indicating and adjustment module PLICSCOM is used for

measured value display, adjustment and diagnosis. It can be

mounted in the following housing versions and instruments:

l

All sensors of the plics

®

instrument family, in the single as

well as in the double chamber housing (optionally in the

electronics or connection compartment)

l

External indicating and adjustment unit VEGADIS 61

Note:

You will find detailed information on the adjustment in the

operating instructions manual of the "Indicating and adjustment

module PLICSCOM".

6.2 Insert the indicating and adjustment module

The indicating and adjustment module can be inserted in the

sensor and removed at any time. It is not necessary to interrupt

the voltage supply.
Proceed as follows:
1 Unscrew the housing cover
2 Place the indicating and adjustment module in the desired

position on the electronics (you can choose any one of four

different positions - each displaced by 90°)

3 Press the indicating and adjustment module onto the

electronics and turn it to the right until it snaps in.

4 Screw housing cover with inspection window tightly back

on

Removal is carried out in reverse order.
The indicating/adjustment module is powered by the sensor,

an additional connection is not necessary.

Function/Configuration

Mounting/dismounting the in-

dicating and adjustment

module