3 exchange the electronics module – VEGA VEGASON 66 Profibus PA User Manual

Page 43

Advertising
background image

?

E017

l

Adjustment span too small

à

Carry out a fresh adjustment and increase the distance

between min. and max. adjustment

?

E036

l

no operable sensor software

à

Carry out software update or send instrument for repair

?

E041

l

Hardware error, electronics defective

à

Exchange instrument or return instrument for repair

?

E113

l

Communication conflict

à

Exchange instrument or return instrument for repair

8.3 Exchange the electronics module

If the electronics module is defective, it can be replaced by the

user.

In Ex applications, only an instrument and an electronics

module with appropriate Ex approval may be used.

If there is no electronics module available on site, one can be

ordered from the VEGA agency serving you.
The order data of the sensor must be downloaded into the new

electronics module. This can be done:

l

at the factory by VEGA

l

or on site by the user

In both cases, the sensor serial number is necessary. The

serial numbers are stated on the type label of the instrument or

on the delivery note.

Sensor serial number

Maintenance and fault rectification

VEGASON 66 - Profibus PA

43

28789

-EN

-060731