In Simatic S7 the measured value is always presented as 0

l

Only four bytes are consistently loaded in the PLC

à

Use function component SFC14 to load 5 bytes

consistently

?

Measured value on the indicating and adjustment module

does not correspond to the value in the PLC

l

The menu item "Display - Display value" is not set to

"PA-Out"

à

Check values and correct, if necessary

?

No connection between PLC and PA network

l

Incorrect adjustment of the bus parameter and the

segment coupler-dependent baud rate

à

Check data and correct, if necessary

?

Instrument does not appear during connection setup

l

Profibus DP cable pole-reversed

à

Check cable and correct, if necessary

l

Incorrect termination

à

Check termination at the beginning and end points of

the bus and terminate, if necessary, according to the

specification

l

Instrument not connected to the segment, double

assignment of an address

à

Check and correct, if necessary

In Ex applications, the regulations for the wiring of intrinsically

safe circuits must be observed.

?

E013

l

no measured value available

à

sensor in boot phase

à

sensor does not find an echo, e.g. because of faulty

installation or incorrect parameter adjustment

Fault messages via the indi-

cating/adjustment module