VEGA VEGASON 66 Profibus PA User Manual
Page 56
Status
code
Description according to Profi-
bus standard
Possible cause
0x44
uncertain - last unstable value
Failsafe replacement value (Failsafe-Mode = "Last value" and already
valid measured value since switching on)
0x48
uncertain substiute set
l
Switch on simulation
l
Failsafe replacement value (Failsafe-Mode = "Fsafe value")
0x4c
uncertain - initial value
Failsafe replacement value (Failsafe-Mode = "Last valid value" and no
valid measured value since switching on)
0x51
uncertain - sensor; conversion not
accurate - low limited
Sensor value < lower limit
0x52
uncertain - sensor; conversion not
accurate - high limited
Sensor value > upper limit
0x80
good (non-cascade) - OK
OK
0x84
good (non-cascade) - active block
alarm
Static revision (FB, TB) changed (10 sec. long active after parameters
of the static category were written)
0x89
good (non-cascade) - active advi-
sory alarm - low limited
Lo-Alarm
0x8a
good (non-cascade) - active advi-
sory alarm - high limited
Hi-Alarm
0x8d
good (non-cascade) - active crit-
ical alarm - low limited
Lo-Lo-Alarm
0x8e
good (non-cascade) - active crit-
ical alarm - high limited
Hi-Hi-Alarm
Supplement
56
VEGASON 66 - Profibus PA
28789
-EN