VEGA SOLITRAC 31 4 … 20 mA_HART - four-wire User Manual

Page 23

Advertising
background image

23

5 Connecting to power supply

SOLITRAC 31 • 4 … 20 mA/HART - four-wire

40090-EN-131119

instructions are part of the scope of delivery and come with the Ex-

approved instruments.

1

1

24

/L

/N

PE

56

22

11

12

21

22

19

20

18
17

14
13

16
15

12

2

5

1

3

4

6

7

8

Fig. 13: Electronics and connection compartment (Ex-d) with instruments with

intrinsically safe current output
1 Voltage supply

2 Relay output

3 Signal input 4 … 20 mA

4 Switching input for NPN transistor

5 Switching input floating

6 Transistor output

7 Interface for sensor-sensor communication (MGC)

8 Setting the bus address for sensor-sensor communication (MGC)

4)

5 6 7 8

2

3

4

1

2

+

( )

(-)

1

Fig. 14: Adjustment and connection compartment (Ex-ia) with instruments with

intrinsically safe current output
1 Terminals for intrinsically safe signal output 8/16 mA/HART (Multidrop) ac-

tive (not on versions with Ex-d approval)

2 Contact pins for the display and adjustment module or interface adapter

3 Terminals for the external display and adjustment unit

4 Ground terminal

Electronics and connec-

tion compartment - In-

struments with intrinsi-

cally safe current output

Adjustment and connec-

tion compartment - In-

struments with intrinsi-

cally safe current output

4)

MGC = Multi Gauge Communication

Advertising