9 Diagnostics and service

SOLITRAC 31 • 4 … 20 mA/HART - four-wire

40090-EN-131119

Maintenance: Due to external influences, the instrument function

is limited. The measurement is affected, but the measured value is

still valid. Plan in maintenance for the instrument because a failure is

expected in the near future (e.g. due to buildup).
This status message is inactive by default. It can be activated by the

user via PACTware/DTM or EDD.

The following table shows the error codes and text messages in the

status message "Failure" and provides information on causes as well

as corrective measures.
Example for a failure message

Code
Text mes-

sage

Cause

Rectification

F008
Error multi

sensor com-

munication

– Additional sensors not

switched on

– EMC influences

– No other sensor available

– Check wiring between the

sensors

– Connect the sensors cor-

rectly and make them ready

for operation

F013
Sensor sig-

nals a fault

– Error on the current input

– No valid measured value

– Connected instruments

without function

– Check current input

– Check connected instru-

ments (Slaves)

F016
Adjustment

data ex-

changed

– Values of the min. and max.

adjustment exchanged

– Correct adjustment data

F017
Adjustment

span too small

– The values of the min. and

max. adjustment are too

close together

– Correct adjustment data

F025
Invalid lineari-

zation table

– Empty linearization table

– Wrong value in the lineari-

zation table

– Create linearization table

– Correct linearization table

F030
Process value

out of limits

– Process values are not

within the adjusted measur-

ing range

– Repeat adjustment

F034
EPROM hard-

ware error

– Electronics defective

– Restart instrument

– Exchanging the electronics

F035
EPROM data

error

– Error in the internal instru-

ment communication

– Carry out a reset

– Exchanging the electronics