VEGA SOLITRAC 31 4 … 20 mA_HART - four-wire User Manual

Page 70

Advertising
background image

70

11 Supplement

SOLITRAC 31 • 4 … 20 mA/HART - four-wire

40090-EN-131119

Process conditions
For the process conditions, please also note the specifications on the type label. The lower value

always applies.
Process pressure

Unpressurized

Process temperature (measured on the

detector tube)

-40 … +60 °C (-40 … +140 °F)
With temperatures of more than 60 °C we recommend

the use of water cooling

Vibration resistance

6)

mechanical vibrations up to 1 g in the frequency range

5 … 200 Hz

Electromechanical data - version IP 66/IP 67
Cable entry

Ʋ M20 x 1.5

2 x cable gland M20 x 1.5 (cable: ø 6 … 12 mm), 4 x

blind plug M20 x 1.5
Included: 1 x cable gland M20 x 1.5

Ʋ ½ NPT

5 x closing cap (red) ½ NPT
Included: 3 x cable gland ½ NPT (cable: ø 6 … 12 mm),

4 x blind plug ½ NPT

Spring-loaded terminals for wire cross-section

Ʋ Massive wire, cord

0.2 … 2.5 mm² (AWG 24 … 14)

Ʋ Stranded wire with end sleeve

0.2 … 1.5 mm² (AWG 24 … 16)

Display and adjustment module
Display element

Display with backlight

Measured value indication

Ʋ Number of digits

5

Ʋ Size of digits

W x H = 7 x 13 mm

Adjustment elements

4 keys

Protection rating

Ʋ unassembled

IP 20

Ʋ mounted into the housing without

cover

IP 40

Materials

Ʋ Housing

ABS

Ʋ Inspection window

Polyester foil

Integrated clock
Date format

Day.Month.Year

Time format

12 h/24 h

Time zone Ex factory

CET

Rate deviation max.

10.5 min/year

6)

Tested according to the guidelines of German Lloyd, GL directive 2.