3 wiring plan – VEGA VEGAWELL 72 4 … 20 mA_HART PT 100 User Manual

Page 20

Advertising
background image

2

I

nsert the connection cable through the cable entry into the

plastic housing

3

L

oosen the screws with a screwdriver

4

I

nsert the wire ends into the open terminals according to

the wiring plan

5

T

ighten the screws with a screwdriver

6

C

heck the hold of the wires in the terminals by lightly

pulling on them

7

T

ighten the compression nut of the cable entry. The seal

ring must completely encircle the cable

8

R

etighten the housing cover

T

he electrical connection is finished.

5

.3 Wiring plan

1

8

2

3

4

5

6

7

Fig. 7: Wire assignment, connection cable
1

blue (-): to power supply or to the processing system

2

brown (+): to power supply or to the processing system

3

white: for processing of the integrated Pt 100 (power supply)

4

Yellow: for processing of the integrated Pt 100 (measurement)

5

Red: for processing of the integrated Pt 100 (measurement)

6

Black: for processing of the integrated Pt 100 (power supply)

7

Shielding

8

Breather capillaries with filter element