Deviation determined according to the limit point method according to IEC 60770

7

)

A

pplies to digital HART interface as well as to analogue current output 4 … 20 mA.

S

pecifications refer to the set span. Turn down (TD) is the relation nominal measuring

range/set span.

D

eviation with version < 0.25 %

-

T

urn down 1 : 1 up to 5 : 1

<

0.25 %

-

T

urn down up to 10 : 1

<

0.25 %

x TD

D

eviation with version < 0.1 %

-

T

urn down 1 : 1 up to 5 : 1

<

0.1 %

-

T

urn down up to 10 : 1

<

0.02 %

x TD

Influence of the product or ambient temperature
A

pplies to digital HART interface as well as to analogue current output 4 … 20 mA.

S

pecifications refer to the set span. Turn down (TD) is the relation nominal measuring

range/set span.

Average temperature coefficient of the zero signal

I

n the compensated temperature range of 0 … +80 °C (+32 … +176 °F), reference

temperature 20 °C (68 °F).

A

verage temperature coefficient of the zero signal

-

T

urn down 1 : 1

<

0.15 %/10 K

-

T

urn down up to 5 : 1

<

0.2 %/10 K

-

T

urn down up to 10 : 1

<

0.25 %/10 K

O

utside the compensated temperature range:

A

verage temperature coefficient of the zero signal

-

T

urn down 1 : 1

typ. < 0.15 %/10 K

Thermal change, current output

A

pplies also to the analogue 4 … 20 mA current output and refers to the set span.

T

hermal change, current output

<

0.15 %

at -40 … +80 °C (-40 … +176 °F)

Long-term stability (similar to DIN 16086, DINV 19259-1 and IEC 60770-1)

A

pplies to digital HART interface as well as to analogue current output 4 … 20 mA.

S

pecifications refer to the set span. Turn down (TD) is the relation nominal measuring

range/set span.

L

ong-term drift of the zero signal

< (0.1 %

x TD)/year

7)

I

ncl. non-linearity, hysteresis and non-repeatability.