7 set instrument address – VEGA VEGAPULS 68 (≥ 2.0.0 - ≥ 4.0.0) Profibus PA User Manual

Page 33

Advertising
background image

33

5 Connecting to the bus system

VEGAPULS 68 • Profibus PA

36537-EN-130205

Contact pin

Colour connection

cable in the sensor

Terminal, electronics

module

Pin 3

Blue

7

Pin 4

Black

8

5.6 Wiring plan - version IP 66/IP 68, 1 bar

1

2

Fig. 35: Wire assignment fix-connected connection cable
1 brown (+) and blue (-) to power supply or to the processing system

2 Shielding

5.7 Set instrument address

An address must be assigned to each Profibus PA instrument. The

approved addresses are between 0 and 126. Each address must

only be assigned once in the Profibus PA network. The sensor is only

recognized by the control system if the address is set correctly.
When the instrument is shipped, address 126 is adjusted. This

address can be used for function test of the instrument and for con-

nection to a Profibus PA network. Then address must be changed to

integrate additional instruments.
The address setting is carried out either via:

The address selection switch in the electronics compartment of

the instrument (address setting via hardware)

The indicating and adjustment module (address setting via soft-

ware)

PACTware/DTM (address setting via software)

The hardware addressing is effective if an address <126 is adjusted

with the address selection switches on the instrument. Hence the

software addressing is no longer effective, the adjusted hardware

address is valid.

Wire assignment, con-

nection cable

Instrument address

Hardware addressing

Advertising