VEGA VEGADIS 62 User Manual

Page 25

Advertising
background image

l

Pressure: bar, mbar, PSI, hPa, kPa, MPa, mmH2O, mH2O, inHg

l

Temperature: °C, °F, K

l

Electrical variables: V, mA, Ohm

l

User-specific unit: USER

The previously listed units can be initially edited. A storing in the
VEGADIS

62 is however only possible if the unit is supported by the

sensor.

Note:
A unit adjusted in the sensor, for example, via PACTware/DTM will not
be modified by the unit adjusted here.

If a unit is selected which is not supported by the sensor, the message
"HART E

rror 7" is displayed. With "ESC" the setting can be interrupted

and the message reset.

Due to the selection of the unit "USER" the HART communication to
the sensor is switched off. Hence no measuring range settings are
possible. The unit "USER" can be individually programmed by the user
in the menu "Configuration". After a reset to a unit supported by the
sensor, the HART communication is again started.

Adjustment of the measuring range initial value of the connected
sensor (for example 0 bar at a measuring range of -1 … 5 bar)

2)

The

measuring range beginning point is initially adopted by the sensor
when starting with connected sensor. Thereafter the measuring range
beginning point of the sensor can be changed.

This is a min. adjustment to which 0 % current output is assigned. The
value entered here does not change the measuring range endpoint.

If the unit is set to "USER", then this menu item is used for free scaling
of the indication.

Adjustment range: -9999 … 99999

2)

Is not available with HART multidrop.

Measurement/Measu-
ring range begin

VEGADIS

62

25