Radar and ultrasonic sensors

l

Sensors with guided microwave

l

Capacitive probes

l

Pressure transmitter

l

Previous instrument versions (replacement for VEGADIS 12)

The operation of respective sensors from other manufacturers is also
possible.

The VEGADIS 62 acts like a HART handheld with limited functions.
The following adjustment functions are available:

l

Min./Max. adjustment

l

zero/span adjustment (live adjustment)

l

Damping

Basic mode: when used in a 4 … 20 mA signal cable, the VEGADIS
62 works as a pure indicating instrument. It measures the current in the
current loop and shows it as digital value as well as via a bargraph. All
settings of VEGADIS 62 are carried out manually via the buttons in the
front.

Adjustment volume: Indication scaling

HART standard: when used with a 4 … 20 mA/HART sensor, the
VEGADIS

62 operates as automatic indicating and HART adjustment

instrument. The VEGADIS 62 powered via the current loop listens
continuously to the HART communication of the control system with
the sensor. Changes of unit and/or measuring range are adapted
automatically.

The parameter adjustment of the sensor is carried out via HART
communication. During the parameter adjustment, the VEGADIS 62
operates as a Secondary Master to the sensor.

Adjustment volume: Sensor functions, indication scaling

HART multidrop: The VEGADIS 62 can also be used as indicator for
a bus participant on a HART multidrop system. For this purpose, it is
looped into the bus and the address of the participant is set in the
VEGADIS

62. The instrument reads out the measured values with unit

via the HART signal and displays them.

Adjustment volume: Sensor function damping, indication scaling

3

.3 Packaging, transport and storage

Your instrument was protected by packaging during transport. Its
capacity to handle normal loads during transport is assured by a test
according to DIN EN 24180.

The packaging of standard instruments consists of environment-
friendly, recyclable cardboard. For special versions, PE foam or PE foil
is also used. Dispose of the packaging material via specialised
recycling companies.