VEGA VEGADIF 65 4 … 20 mA_HART User Manual

Page 11

Advertising
background image

11

3 Product description

VEGADIF 65 • 4 … 20 mA/HART

36128-EN-130417

A metallic measuring cell is used as sensor element. The process

pressures are transmitted via the separating diaphragms and filling

oils to a resistance measuring bridge (semi-conductor technology).
The difference between the acting pressures generates a change in

the bridge voltage. This change is measured, further processed and

converted into a corresponding output signal.
The markings "+" and "-" on the process component in chapter

"Mounting and connection instructions" must therefore be noted when

connecting to the process. The pressure acting on "+" goes positive,

the pressure acting on "-" goes negative into the calculation of the

pressure difference.
The configuration of the measuring cells differs depending on the

measuring range:

1
2

3

4

5

1

p

2

p

Fig. 8: Metallic measuring cell 10 mbar and 30 mbar - p

1

and p

2

process pres-

sures
1 Measuring element

2 Silicone diaphragm

3 Separating diaphragm

4 Filling oil

5 Integrated overvoltage arrester

1

p

2

p

1

2

3

4

Fig. 9: Metallic measuring cell from 100 mbar - p

1

and p

2

process pressures

1 Measuring element

2 Overload diaphragm/Middle diaphragm

3 Filling oil

4 Separating diaphragm

4 … 20 mA/HART two-wire electronics for voltage supply and meas-

ured value transmission on the same cable.
The supply voltage range can differ depending on the instrument

version.
The data for power supply are specified in chapter "Technical data".

Functional principle

Voltage supply

Advertising