5 accessories and replacement parts – VEGA VEGADIF 65 4 … 20 mA_HART User Manual

Page 13

Advertising
background image

13

3 Product description

VEGADIF 65 • 4 … 20 mA/HART

36128-EN-130417

Storage and transport temperature see chapter "Supplement -

Technical data - Ambient conditions"

Relative humidity 20 … 85 %

3.5 Accessories and replacement parts

The display and adjustment module PLICSCOM is used for measured

value indication, adjustment and diagnosis. It can be inserted into the

sensor and removed at any time.
You can find further information in the operating instructions "Display

and adjustment module PLICSCOM" (Document-ID 27835).

The interface adapter VEGACONNECT 4 enables the connection of

communication-capable instruments to the USB interface of a PC. For

parameter adjustment of these instruments, an adjustment software

such as PACTware with VEGA-DTM is required.
You can find further information in the operating instructions "Interface

adapter VEGACONNECT" (Document-ID 32628).

VEGADIS 61 is suitable for external measured value indication and

adjustment of plics® sensors. It is connected to the sensor with an up

to 25 m long, four-wire, screened standard cable.
You can find further information in the operating instructions "VE-

GADIS 61" (Document-ID 27720).

VEGADIS 62 is suitable for measured value indication and adjustment

of sensors with HART protocol. It is looped into the 4 … 20 mA/HART

signal cable.
VEGADIS 62 is suitable for measured value indication with sensors

without HART protocol.
You can find further information in the operating instructions "VE-

GADIS 62" (Document-ID 36469).

The protective cover protects the sensor housing against soiling and

intense heat from solar radiation.
You will find additional information in the supplementary instructions

manual "Protective cover" (Document-ID 34296).

The oval flange adapter enables the connection of a ½ NPT tube to a

VEGADIF 65 or a valve block. By selecting the suitable materials, the

oval flange adapter can be adapted to all processes.
You can find additional information in the supplementary instructions

"Mounting accessory pressure" (Document-ID 43478).

Valve blocks enable simple installation and setup of a differential

pressure transmitter. When the process valves are closed, the equali-

zation valve enables a pressure compensation for the measuring

chambers. This allows the zero point of the pressure transmitter to be

adjusted.

Storage and transport

temperature

Display and adjustment

module

Interface adapter

External display and

adjustment unit

External display and ad-

justment unit VEGADIS 62

Protective cap

Oval flange adapter

Valve blocks

Advertising