6 Adjustment with the display and adjustment module PLICSCOM

VEGADIF 65 • 4 … 20 mA/HART

36128-EN-130417

Part sum counter
0 % = +0000
m

3

/s=

100 % = +0000
m

3

/s=

6. Edit with [OK] and set the requested values with [+] and [->].
7. Confirm with [OK] and jump back to the indication of the part sum

counter.

8. Select with [->] the unit of the sum counter, adjust the requested

unit with [->], for example m

3

/s and confirm with [OK].

The setting of the part sum counter is hence terminated, the counting

function is activated.
The procedure of the total sum counter is the same.

This function enables uploading parameter adjustment data into the

display and adjustment module as well as downloading parameter

adjustment data into the sensor. A detailed description of the function

is available in the operating instructions manual "Display and adjust-

ment module".
The following data are loaded or downloaded with this function:

Measured value presentation

Applications

Adjustment

Damping

Linearization curve

Leak flow volume suppression

Sensor-TAG

Displayed value

Display unit

Scaling

Current output

Unit of measurement

Language

The following safety-relevant data are not uploaded or downloaded:

HART mode

PIN

Copy sensor data

Copy sensor data?

Basic adjustment

The reset "Basic adjustment" resets the values of the following menu

items to the reset values (see chart):

Copy sensor data

Reset

Advertising