2 specifications – YSI Level Scout User Manual

Page 42

Advertising
background image

40

Level Scout

YSI Incorporated

5.2 SPECIFICATIONS

These specifications are subject to change without notice. For the most recent specifications please visit www.ysi.com.


Level

Range:

0 - 10 ft, 3.05 m, 4.3352 PSI

(Vented) 0

-

50

ft,

15.24

m,

21.676

PSI

0 - 75 ft, 22.86 m, 32.5140 PSI

0

-

250

ft,

76.2

m,

108.38

PSI

0

-

692

ft,

210.6

m,

299.9956

PSI

Range:

0-10

ft.

3.05

m,

19.035

PSI

(Absolute)

0-50

ft,

15.24

m,

36.376

PSI

0-70

ft,

21.33

m,

45.046

PSI

0-230

ft,

70.1

m,

114.410

PSI

0-692

ft,

210.9

m,

314.696

PSI

Accuracy:

0.05 % FS TEB* for ranges > 10 ft
± 0.10 % FS TEB* for ranges <= 10 ft

Resolution: 0.00001 (Data Scout and Data Scout Mobile)

0.000001 (Data Scout Advanced, user selectable)

Units:

psi, ft H

2

O, mm H

2

O, cm H

2

O, m H

2

O, kPA

Temperature

Range:

-20 to 50° C (Operating Range)

-10 to 40° C (Compensating Range)

Accuracy:

± 1.0° C (for S2, T2, AS2, AT2 models)

± 0.2° C (for S4, T4, AS4, AT4 models)

Resolution:

0.001

(Data

Scout

and

Data

Scout

Mobile)

0.01 (Data Scout Advanced)

BaroScout Range:

8-16

psia,

55-110

kPa

Modes

of

Sampling

Linear, Linear Average, Event, and Logarithmic

Sampling Rate

Programmable, 8 readings per second max

Memory

2 MB (for S2, T2, AS2, AT2 models)

4 MB (for S4, T4, AS4, AT4 models)


Maximum Pressure Scans

196,560 (for S2, T2, AS2, AT2 models)

(Pressure with Time-Stamp)

393,120 (for S4, T4, AS4, AT4 models)


Maximum Pressure and

144,144 (for S2, T2, AS2, AT2 models)

Temperature Scans

288,288 (for S4, T4, AS4, AT4 models)

(Pressure and Temp with Time Stamp)

Time-Stamp Accuracy

±2 min/year (over the compensated temp range)

Proof Pressure

2 x FS

Burst Pressure

3 x FS

Data Upload Time 100

scans/sec

Storage

Temperature

-40 to 80° C without batteries, -20 to 50° C with batteries

Wetted

Material

316 SS or Titanium; Viton®

Protection Rating

IP 69, NEMA 6P

Internal

Batteries

(2) 1.5 volt AA Alkaline RECOMMENDED

Battery Life

3 years @ 15 minute sample intervals

External Excitation

6

16

VDC

External Input Current

8.0 mA (average current during measurement)

25.0 mA (15 mS peak current during page writes)
0.25 mA (sleep current)

Advertising