YSI Level Scout User Manual

Page 52

Advertising
background image

50

Level Scout

YSI Incorporated

issued.

Example: 0D0! (Send Get Data
command to sensor “0))

would be “avu” form and would indicate
that the water level of sensor “0” is a
10.23 feet.

Note: If address is returned alone with
no data and unit values, this indicates
that there is no data available. Either a
Start Measurement command was not
issued, the command was aborted by
sending a new command before the
measurement time expired, or an
expected Service Request response was
not yet received.

M

or M0

Start non concurrent
pressure measurement

aM!

:

aD0!

attt2

:

a (service request response)

“2”) is number values in “D0” resp:

apu or aputu

“p” is ± pressure value

“u” is + units indicator. Use “XUP”
command to set units. If “XUM=1”
command is executed then ±
temperature value and units code (tu)
also appended.

M1

Start non concurrent
factory pressure
measurement

(factory calibrated
value always in psi)

aM1!

:

aD0!

attt1

:

a (service request response)

“1” is number values in “D0” resp:

ap

“p” is ± pressure value (psi)

M2

Start non concurrent
temperature
measurement

aM2!

:

aD0!

attt2

:

a (service request response)

“2” is number values in “D0” resp:

atu

“t” is signed temperature, “u” is = units:
0-Celsius, 1=Fahrenheit. Use “XUT”
command to set units.

M3

Start non concurrent
scale factor
measurement

(user slope, user offset,
and field calibration

aM3!

:

aD0!

a0013

:

a (service request response)