Performance verification procedure, Vvin vout gain dc – B&K Precision 2542B-GEN - Manual User Manual

Page 240

Advertising
background image

Performance Verification Procedure

240

8. Connect calibrator to oscilloscope. If preferred calibrator is

not available, connect alternative equipment as follows:

9. Apply a reference signal. The output level of the DC

positive/negative of calibrator output should be equal to 3
times the volts/div setting of oscilloscope. For example, to
test 10 mV/div in CH1, the output of the calibrator should be
set to +30 mV/-30 mV.

10.

Compare the reading of the Vavg value at the bottom of the

screen (real time reading of the input signal) to the amplitude of
your reference signal.

11.

The DC gain should always be ≤ 4% for 2 to 5 mV/div and ≤ 3%
for 10 mV to 5 V/div.

+

+

=

=

DMM

DMM

pe

oscillosco

pe

oscillosco

V

V

V

V

Vin

Vout

Gain

DC

In above example, the difference between positive and
negative input value is 60 mV.

12. Select the next volts/div setting.
13. Repeat the above steps for channel 2.

Oscilloscope

DC Power

Supply

DMM

Advertising