Getting Started

29

probe for the first time to any input channel. A poorly

compensated probe can introduce measurement errors.

1. Set both the probe and the oscilloscope attenuation

factor to X10 respectively.

2. Connect the oscilloscope probe to channel 1. Attach the

probe tip and reference lead to the probe compensation

terminal and to the chassis ground terminal, then press

AUTO key.

3. Use a nonmetallic tool to adjust the trimmer capacitor on

the probe for the flattest pulse possible (see “Correct

compensation” image below). The trimmer capacitor is

located either on the probe BNC connector or above the

probe attenuation switch.

Correct compensation

Over compensated

Under compensated

4. Connect probes to channel 2. Repeat the above steps.

This matches each probe to each channel.